Consider the following:
Image acquisition
Image acquisition is obtained with a low frequency curved probe, usually at setting of around 3-5 mHz.
Depth
It is generally advised that the depth is initially set to 16 cm (see bottom right of the image), and reduced for a thinner patient or increased for a larger patient.
Time Gain Compensation (TGC)
If your machine has Time Gain Compensation (TGC), set it with less gain in the near fields and proportionally more in the far fields.
This image shows TGC set with less gain in the near fields and proportionally more in the far fields.
Label the image
Label the image with appropriate patient details. These will usually appear at the top of the scan.
Pictogram
Always add either a pictogram or a text description of what the image is intended to be. A pictogram is a small body picture (see bottom right of the image). Some operators prefer a simple text insert instead, such as ‘RUQ’.
For more information on this topic, see the session about image acquisition.